QUEZADA JACOME, Diana Fernanda; LINDEN, Rafael; GRANDO, Ana Paula y PACHAR, Samantha. Analytical method for vancomycin determination in human plasma through HPLC-UV, using LC-MS/MS data comparison for its validation. Rev.Cs.Farm. y Bioq [online]. 2025, vol.13, n.2, pp.15-26. Epub 08-Abr-2026. ISSN 2310-0265. https://doi.org/10.53287/bfoi3796ig50o.
Objective.
To valídate a vancomycin monitoring assay through HPLC-UV for its clinical therapeutic application.
Materials and methods.
Vancomycin was extracted from human plasma by protein precipitation using perchloric acid and dichloromethane wash. HPLC-UV separation was performed on a reverse-phase column with isocratic elution and UV detection. For method comparison, a human plasma was obtained from a reference laboratory and the results were analysed using Medícale statistical software.
Results.
The HPLC-UV analysis time was 22 minutes. The assay was linear in the range of 2.5 to 100 mcg/mL; the aecuracy was 96.8 -113.6 %, the intra-assay precisión was 4.65 - 9.30% and, the inter-assay precisión was 6.84 - 7.50%. The assay was applied on 40 human plasma samples, the results were proportional and showed a positive correlation with no systematic differences to the determined by LC-MS/MS reference method.
Conclusion.
The vancomycin detection assay using HPLC-UV was validated, showing high correlation with highly sensitive techniques (LC-MS/MS). The method proved to be selective, sensitive, and easy to implement in clinical laboratories with the aim of pharmacokinetically individualizing vancomycin doses.
Palabras clave : HPLC-UV; LC-MS/MS; vancomycin; therapeutic drug monitoring (TDM).